久久无码人妻一区二区三区午夜_久久久久精品久久久久影院蜜桃_亚洲综合欧美色五月俺也去_交换娇妻呻吟声不停中文字幕

產品展廳收藏該商鋪

您好 登錄 注冊

當前位置:
世聯博研(北京)科技有限公司>>微電極>>微電極>>MI-411 MI-413針式組合PH微電極

針式組合PH微電極

返回列表頁
  • 針式組合PH微電極

收藏
舉報
參考價 面議
具體成交價以合同協議為準
  • 型號 MI-411 MI-413
  • 品牌 其他品牌
  • 廠商性質 代理商
  • 所在地 北京市

在線詢價 收藏產品 加入對比 查看聯系電話

更新時間:2017-12-29 15:46:30瀏覽次數:723

聯系我們時請說明是化工儀器網上看到的信息,謝謝!

產品簡介

供貨周期 現貨    
針式組合PH微電極
The microelectrode is ready to use. Carefully unwind the tape and remove the probe from the
protective glass tube. The white translucent sleeve, which covers the hole, should be moved
down

詳細介紹

針式組合PH微電極

Calibration
The electrode can be held by hand when measuring pH. It has been standardized in two (2) pHbuffers. Follow the procedure recommended by the manufacturer of your pH meter forcalibration.

針式組合PH微電極

Optimum Response Time: Optimum response time will be obtained after the probe has been
exercised in two (2) buffer solutions. Fill one beaker with a pH 4.01 buffer or equivalent and a
second beaker with a pH 6.86 buffer or equivalent. Hold the microelectrode and touch the pH 4
.01 buffer surface allowing 5 seconds for equilibration, and then touch the pH 6.86 buffer surface
in the same0 manner. Do this several times. It is not necessary to rinse the probe with distilled
water because the carry-over is negligible

Handling
The outer reference chamber of the microelectrode can be refilled when necessary with the 3 M
KCl solution, using the bulb pipette. Hold the electrode at a 60 degree angle (approx.) and
slowly insert the tip into the fill hole of the electrode.
Be careful not to apply pressure against the inner glass capillary tube.
Slowly fill the reference chamber. If filling is done too quickly, air may be trapped in the
chamber causing reference solution to squirt out of the fill hole instead of entering and filling the
reference chamber.
Cleaning
When using the microelectrode in solutions containing protein, it should be soaked in an enzyme
cleaning solution such as Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning
solution for a couple of minutes after each use to remove the protein from the glass and the
reference junction. This will prolong the useful life of the microelectrode

收藏該商鋪

登錄 后再收藏

提示

您的留言已提交成功!我們將在第一時間回復您~

對比框

產品對比 產品對比 聯系電話 二維碼 意見反饋 在線交流

掃一掃訪問手機商鋪
86-010-67529703
在線留言
主站蜘蛛池模板: 丰原市| 巴林右旗| 宁陵县| 安顺市| 洪湖市| 卓尼县| 洛南县| 萨迦县| 鄂温| 谢通门县| 道真| 日喀则市| 曲靖市| 和林格尔县| 新宁县| 安陆市| 扎鲁特旗| 寿宁县| 华池县| 旌德县| 南丹县| 富蕴县| 清苑县| 嘉义县| 尼木县| 儋州市| 台安县| 安溪县| 安丘市| 红安县| 阳城县| 汉阴县| 洞头县| 临西县| 大化| 金湖县| 慈溪市| 云霄县| 尉氏县| 沾益县| 临泽县|